Zimmer ATS 750
Portable tourniquet to temporarily occlude blood flow in a patient’s extremities during surgical procedures on those extremities.  A tool to produce a bloodless operation field in surgical procedures involving the extremities including:
Reduction of certain fractures     Kirschner wire removal     Tumor and cyst excisions     Subcutaneous fasciotomy     Nerve injuries    Tendon repair     Bone grafts     Total wrist joint replacement     Replacement of joints of the fingers     Knee joint replacements    Amputations     Replantations
- Description
- Specification
- Documents
Description
Zimmer ATS 750 DESCRIPTION
The Zimmer A.T.S. 750 Tourniquet is a full-featured and automatic tourniquet system that is sensitive to the cost issues that all health care professionals face today. Its features include:
- Uses ambient air
- Uses electronic and microprocessor regulated pressure
- Single port system: Uses single port, single bladder cuffs
- 30 minute battery back-up provides mobility by helping ensure continuous power
- Visual and audible alarms activated when an unsafe condition is detected
- Self-check calibration each time the A.T.S. 750 System is switched on
SPECIFICATIONS
General
- Overall Dimensions: 8″ (20.3 cm) H x 7″ (17.8 cm) W x 7.9″ (20.1 cm) D
- Weight: 7.7 lbs. (3.5 kg)
Timer
- Time Alarm Set Ranges: 5 – 240 minutes; 5 minutes increments
- Timer Accuracy: 0.25% of elapsed time
Power
- Line Voltage Range: 100 – 240 VAC, 50/60 Hz., Auto Switching
- Line Current: 670 mA RMS @ 120 VAC
- Input Power: 53 Watts Typical
- Power Cord: Type SJT, AWG 16, 14 ft. (4.27 m)
- Power Plug: Hospital grade, 3 prong straight blade, 15 Amp
Pressure
- Cuff Pressure Range: 50 – 475 mmHg, 5 mmHg increments
- Pressure Accuracy: ±5 mmHg (50 – 475 mmHg)
- Pressure Regulation: ±6 mmHg of set point (10 second average under non-transient conditions without external leaks)
- Maximum Pressures: 475 mmHg cuffs
Battery
- Battery Type: Rechargeable, 12VDC sealed lead acid, 1.2 Amp hours
- Battery Discharge Time: Unit will operate on battery power for 30 minutes minimum with fully charge batteries
- Battery Recharge Time: 24 hours (Maximum) Unit should be plugged in 24 hours before initial use.
Additional information
| Weight | 7.7 lbs | 
|---|---|
| Dimensions | 8 × 7 × 8 in | 
| Part Number | 60-0750-101-00 | 
| Features | |
| Channels | |
| Cuff Port | |
| Pressure Range | |
| Accuracy | |
| Battery Run Time | |
| Battery Chemistry | Sealed Lead Acid | 
 
								
 
								 
								
 