Draeger Fabius GS
Type: Anesthesia Machine
Make: Draeger
Model: Fabius GS
Condition: Refurbished
Availability: In-Stock
Status: Supported by Denova Medical
- Description
- Specification
- Documents
Description
Dräger Fabius GS Features:
- Proven Ventilation Versatility
- Electronic Gas-Flow Measurement
- Compact, Convenient Breathing System
- Intelligent Cable Management
Specifications
Weight:
- 224 lbs. (101.6 kg) (base unit without vaporizers or cylinders.)
Dimensions:
- (W) 89.5 cm x (H) 130 cm x (D) 82 cm (35.2 x 51.2 x 32.3 inches)
Power Supply:
- 100 – 240 VAC, 50/60 Hz, 2.3 A max.
- Battery (supports ventilator and monitor) > 45 min.
Ventilator:
- E-vent Electronically controlled, electrically driven.
Operating Modes Standard:
- Manual/Spontaneous.
- Volume Control (IPPV).
Options:
- Pressure Control. (PCV)
- Pressure Support. (PS)
- Synchronized Volume Controlled Ventilation w/PS. (SIMV/PS)
Breathing Frequency:
- 4 to 60 bpm.
Max. Minute Volume (MV):
- 25 L/min.
PEEP:
- 0 – 20 cmH2O.
Inspiration / Expiration Ratio:
- (Ti:Te) 4 : 1 to 1 : 4.
Pressure Limiting (Pmax):
- 15 – 70 cmH2O.
Tidal Volume (Vt):
- 20 – 1400 mL in Volume Control.
- 20 – 1100 mL in SIMV/PS.
Inspiratory Pause (Tip:Ti):
- 0 – 50 .
SIMV Inspiratory Time(Tinsp):
- 0.3 – 4.0 sec.
Inspiratory Pressure (Pinsp):
- PEEP + 5 to 65 cmH2O.
Inspiratory flow (InspFlow):
- 10 – 75 L/min in Volume and Pressure Control.
- 10 – 85 L/min in Pressure Support.
Pressure Support Level (Δ PPS) PEEP:
- + 3 to 20 cmH2O Min. Frequency for 3 – 20 bpm and “ OFF ”.
Apnea-Ventilation:
- ( Freq. Min.)Trigger 2 – 15 L/min.
Safety functions:
- Sensitive Oxygen Ratio Controller (S-ORC) guarantees a minimum O2 concentration of 23% in an O2/N2O mixture. N2O cut-off if O2 fresh gas valve is closed or if O2 flow is less than 0.2 L/min.
Safety functions:
- Sensitive Oxygen Ratio Controller (S-ORC) guarantees a minimum O2 concentration of 23% in an O2/N2O mixture. N2O cut-off if O2 fresh gas valve is closed or if O2 flow is less than 0.2 L/min.
- Audible and visual (flashing red LED)indication in case O2 pressure drops below 20 psi (1.38 bar) ± 4 psi (0.27 bar). In case of electricity and battery failure, manual ventilation, gas delivery and gent delivery are possible. Positive pressure relief valve opens at 75 ± 5 cmH2O. Negative pressure relief valve opens at – 7.5 to – 9 cmH2O.
Range of fresh gas flow indicators:
- 0.00 to 12.0 L/min.
Total fresh gas flow meter:
- 0 to 10 L/min, calibrated with a mixture of 50 % O2 and 50 % N20 mixture.
O2 flush at 87 psi (6 bar):
- max. 75 L/minat 41 psi (2.8 bar): min 25 L/min.
Vaporizer Mount:
- Dräger or SelectatecR mount.
Monitoring:
- Continues monitoring of inspiratory O2 concentration (can be switched off by Service) breathing frequency, tidal volume, minute volume, mean or plateau pressure, peak airway pressure as well< as PEEP. In addition, all fresh gas flow information is displayed as virtual flow tubes.
Serial interface:
- 1 x RS 232 (standard); 1 x RS 232. (option)
Additional information
Manufacturer | |
---|---|
Name | |
Part Number | Fabius GS |
Specialty | |
Application | |
Product Type | |
Ventilator Modes | |
Ventilator Type | |
Vaporizer mounts | |
Battery Run Time |