Zimmer ATS 1200
Type: Tourniquet
Make: Zimmer
Model: A.T.S. 1200
Condition: Refurbished
Availability: In-Stock
Status: Supported by OEM
- Description
- Specification
- Documents
Description
Zimmer ATS 1200 DESCRIPTION
The lightweight Zimmer A.T.S. 1200 Tourniquet System is a single-port, dual-cuff surgical tourniquet system with microprocessor controls and proprietary safety features that make tourniquet application and monitoring convenient and reliable. These key design elements free the healthcare team to focus on surgery. Its features include:
- Uses ambient air
- Uses electronic and microprocessor regulated pressure
- Single port, dual cuff system: Uses up to two single port, single bladder cuffs
- Allows for use of IVRA/Bier Block procedures
- 45 minute battery back-up provides mobility by helping ensure continuous power
- Visual and audible alarms activated when an unsafe condition is detected
- Self-check calibration each time the A.T.S.1200 System is switched on
- Lightweight, portable unit can be mounted on an I.V. pole or placed on a table
SPECIFICATIONS
General
- Overall Dimensions: 12.50″ (31.75 cm) H x 10.25″ (26 cm) W x 8.1″ (20.6 cm) D
- Weight: 8.0 lbs. (3.63 kg)
Timer
- Time Alarm Set Ranges: 5 – 240 minutes; 5 minutes increments
- Timer Accuracy: 0.25% of elapsed time
Power
- Line Voltage Range: 100 – 240 VAC, 50/60 Hz., Auto Switching
- Line Current: 670 mA RMS @ 120 VAC
- Input Power: 53 Watts Typical
- Power Cord: Type SJT, AWG 16, 14 ft. (4.27 m)
- Power Plug: Hospital grade, 3 prong straight blade, 15 Amp
Pressure
- Cuff Pressure Range: 50 – 475 mmHg, 5 mmHg increments
- Pressure Accuracy: ±5 mmHg (50 – 475 mmHg)
- Pressure Regulation: ±6 mmHg of set point (10 second average under non-transient conditions without external leaks)
- Maximum Pressures: 475 mmHg cuffs
Battery
- Battery Type: Rechargeable, 12VDC sealed lead acid, 2.3 Amp hours
- Battery Discharge Time: Unit will operate on battery power for 45 minutes minimum with fully charge batteries
- Battery Recharge Time: 24 hours (Maximum) Unit should be plugged in 24 hours before initial use.
Additional information
Manufacturer | |
---|---|
Name | |
Part Number | 60-1200-101-00, 60120010100 |
Specialty | |
Cuff Port | |
Product Type | |
Channels | |
Cuff Size Range | |
hrs | |
Features |