Type: Electrosurgical Unit
Make: Conmed
Model: Saber 2400, 60-5600-002
Condition: Refurbished
Availability: In-Stock
Status: Supported by Denova Medical, Inc.
- Description
- Specification
- Documents
Description
Conmed Sabre 2400 Specifications
General
- Pure Cut
- Max Power: 300 W.
- Max. P-P Voltage: 2000<./li>
- Load (ohms): 300.
- Crest Factor: 1.8.
- Blend Cut
- Max Power: 180 W.
- Max. P-P Voltage: 2200.
- Load (ohms): 300.
- Crest Factor: 2.5.
- Coag
- Max Power: 120 W.
- Max. P-P Voltage: 9000.
- Load (ohms): 500.
- Crest Factor: 7.0.
- Bipolar
- Max Power: 50 W.
- Max. P-P Voltage: 280.
- Load (ohms): 50.
- Crest Factor: 1.5 – 12.1.
Dispersive Electrode Monitor
- With a dual foil dispersive electrode, the Aspen Return Monitor will alarm above 150 ohms, below 10 ohms, and with a 20% increase in resistance above set point. With a single foil dispersive electrode, a two-wire continuity detector alarms at 10 ohms and above.
Operating Frequency
- Cut: 416.7 kHz ± 0.01%.
- Coag: 416.7 kHz ± 0.01%.
Pulse Repetition Frequency
- Blend: 20.0 kHz ± 0.01%.
- Coag: 25.0 kHz ± 0.01%.
- Bipolar: 20.0 kHz ± 0.01%.
Input Voltage
- 104-127 VRMS, 50/60 Hz, calibrated at 120V.
Input Current
- Pure cut at 300: 7.0A max.
Line Frequency Leakage.
- Patient connections to neutral: < 25 μA.
- Chassis to neutral: < 25 μA typical.
RF Leakage Current
- Less than 100 mA (200 ohms to ground).
Power Decrease
- 1% typical. Pure cut set to 150 watts for 20 seconds.
Weight
- 32 lbs (14.5 kg).
Dimensions
- 6.75″H x 14.25″W x 21.5″D (17.1 x 36.2 x 54.6 cm).
Additional information
Manufacturer | |
---|---|
Name | |
Part Number | 60-5600-002 |
Specialty | |
Maximum Cut Power | |
Product Type | |
Cut Frequency | |
Cut Modes | |
Maximum_Coagulation_Power | |
Coag Frequency kHz | |
Coagulation Modes | |
Bipolar Output Power | |
Bipolar Frequency | |
Type | |
Return Electrode Monitoring |